SeeLens HP

Hydrophobic Aspheric IOL. Providing Aberration Free Vision

Technical Specifications

Overall Length………… 13.00 mm
Optic diameter…………… 6.00 mm
Haptic angulation………..
   
Optic Design……………. Aspheric bi convex
Edge Design……………. 360° Continuous Square Edge
Power range…………….. +5.0 to +10.0 (1D increments)
+10.5 to +30.0 (0.5D increments)
+31.0 to +35.0 (1D increments)
Material…………………… Hydrophobic Acrylic incorporates a natural yellow filtering
Refractive Index…………. 1.484 (@ 35° c)
Y.A.G laser……………….. Compatible
A constant (SRK/T) … Optical / Immersion US biometry: 119.5*
Contact US biometry: 119.16*
Placement………………… Capsular Bag
CE Approved
 

*It is recommended that surgeons personalize their A-constant based on their surgical techniques and equipment, experience and post-operative results. For more information please visit Hanita Lenses web site

 

Attributes

Advanced Optical Design


The Aspheric SeeLens HP was designed using the most advanced tools by a professional R&D team of optical and mechanical engineers.
The optical profile of the SeeLens HP was calculated using ZEMAX™ software – a simulating tool for the optical design optimization. Calculations were aimed to minimize all aberrations, including the spherical aberration of the cornea, and to optimize the MTF (Modulated Transfer Function).

 

Eye Model


The Optical design of SeeLens HP was performed using the Arizona Eye Model[1]. The parameters and dimensions of the eye model are consistent with average human data. The model was designed to match clinical levels of aberration, both on and off axis. The retina curvature is designed to split the tangential and sagittal foci off-axis. The result is an accurate simulation of the visual performance of the SeeLens HP in a post-operative eye.

 

[1] Field Guide to Visual and Ophthalmic Optics; Jim Schwiegerling; Nov. 2004

 

Geometrical Design


  • SeeLens HP ensures excellent stability and centration due to the unique C-loop mechanical design, ensuring wide angle of contact.
  • 360° square edge in order to minimize PCO.

 

Material

 

The SeeLens HP is made of a hydrophobic material with bio-adhesive characteristics, resulting in reduction of secondary cataract procedures [1].
The SeeLens HP has a very a very low glistening effect [2 3].
The SeeLens HP hydrophobic material incorporates a natural yellow, violet filtering, chromophore for better protection of the retina.

 

[1] Advances in Intraocular Lens Material and Designs: Maximizing Biocompatibility and Optical Performance; Ophthalmogica 2003;217
[2] Data from Benz Rechearch and Development QA Dept
[3] Glistening in the Single -Piece Hydrophobic, Acrylic Intraocular Lenses; Aaron Waite, Nathan Faulker and Randall J Olson; J Ophthalmol 2007; 144: 143- 144

 

Mini Incision implantation


  • Easy and safe injection through an incision of 2.4 mm
  • Excellent memory – superior foldability,  slow release and gentle unfolding

 