Technical Specifications
| Overall diameter………. | 11.0 mm |
| Optic diameter…………… | 6.00 mm |
| Standard Power Range | |
|
Power (spherical equivalent): Cylinders |
+10.0D to +30.0D (0.5D increments) +1.0D to +6.0D (0.5D increments) |
| Additional Toric powers can be supplied within the range: | |
|
Power (spherical equivalent): Cylinders |
+30.5D to +40.0D (0.5D increments) .+6.5D to +10.0D (0.5D increments) |
| Optic design……………… | Aspheric biconvex design with Toric surface on anterior plane |
| Edge design……….. | 360° square edge |
| Haptic angulation……….. | 5° |
| Material…………………… | Hydrophilic Acrylic HEMA/EOEMA copolymer |
| Refractive Index…………. | .1.46 (hydrated @ 35°c) |
| Y.A.G laser……………….. | Compatible |
| Estimated A constant… |
118.5 * SRK/T IOL Master Biometry 118.2 * SRK/T US Biometry |
| Placement………………… | Capsular Bag |
| CE Approved | |
* It is recommended that surgeons personalize their A-constant based on their surgical techniques and equipment, experience and post-operative results. For more information please visit Hanita Lenses web
site
Attributes
Advanced Optical Design
Eye Model
The parameters and dimensions of the eye model are consistent with average human data. The model was designed to match clinical levels of aberrations, both on and off axis. The retina curvature is designed to split the tangential and sagittal foci off-axis, providing an optimal platform for correction of toric aberration. The result is an accurate simulation of the visual performance of the BunnyLens TR in the Postoperative eye.
]1 ] Field Guide to Visual and Ophthalmic Optics; Jim Schwiegerling; Nov 2004
Geometrical Design
- BunnyLens TR ensures excellent stability and centration due to the 4-point haptic design
- 360° square edge in order to minimize PCO
Hanita Lenses Toric IOL Calculator
The preoperative parameters include the incision location and the Surgical Induced Astigmatism both supplied by the surgeon
- Flat Keratometry (in D) …………….. 40
- Flat axis ………………………………….. 20
- Steep Keratometry (in D) …………… 45
- Steep axis ……………………………….. 110
- Axial Length ……………………………. 21
- Surgically Induced Astigmatism …. 0.5
- Incision Location (IL) ……………….. 90
Placement of the IOL is provided by the experienced team of engineers in Hanita Lenses based on optical calculations
Stability and Centration
High level of rotational stability is maintained due to the wide angle of contact with the capsular bag.
Material
- The BunnyLens TR is made of a hydrophilic acrylic material, with proven reputation and many years of clinical experience
- The BunnyLens TR is characterized by excellent biocompatibility and mechanical quality
- The BunnyLens TR material incorporates a violet filtering chromophore for better protection of the retina