Short-Term Visual Outcomes After Bilateral Pentafocal Intraocular Lens Implantation. A Pilot Study

Kepa Balparda, MSc PhD, Laura Zharick Ortega-Quiroga, MD, Mariana Escobar-Giraldo, MD, Daniela Ortiz-Piedrahíta, Isabela Franco-Sánchez, Tatiana Herrera-Chalarca, BSc.
(2024)

This study aimed to assess the short-term refractive and visual outcomes following the implantation of the Hanita Intensity SL pentafocal intraocular lens (IOL) in patients with presbyopia. In a retrospective observational review, 20 eyes from 10 patients (70% female) who underwent immediate sequential bilateral surgery were evaluated over three months. Results showed that 100% of patients achieved visual acuity of 20/25 or better for both near and far distances, and 20/40 or better for intermediate distances. The defocus curve indicated excellent visual performance from infinity to approximately 33 cm, allowing functional vision at all distances. Additionally, all patients reported complete independence from glasses. These findings suggest that the Hanita Intensity SL IOL is a safe and effective option for achieving spectacle independence in presbyopic patients.

Objective:

To evaluate short-term refractive and visual behavior after implantation of a new pentafocal
intraocular lens to manage presbyopia in otherwise healthy patients.

Methodology:

A retrospective observational study was conducted, in which the postoperative outcome
was evaluated three months after surgery. Patients in whom the Hanita Intensity SL intraocular lens was
implanted in immediate sequential bilateral surgery were assessed. Medical records were reviewed to
collect information on corrected and uncorrected vision, subjective refraction, and defocus curves with
correction.

Results:

Twenty eyes belonging to 10 patients were included. There was a preference for the female sex
(70%). The 100% of patients achieved vision equal to or better than 20/25 in near and far vision; 100%
of the patients achieved a vision equal to or better than 20/40 in intermediate vision. The defocus curve
showed excellent visual performance in the vergences from infinity to approximately 33cm, with functional
vision at all distances; 100% of the patients reported being free of glasses for all distances.

Conclusion:

Hanita Intensity SL intraocular lens implantation is a safe and effective technique for managing
presbyopia. In well-selected patients, it can provide a high proportion of independence in spectacle wear
for distances up to approximately 33cm.

STAY INFORMED

Signup to get the first look at new products, publications, and events

Related publications

SRYAHWA PUBLICATIONS

Archives of Ophthalmology and Optometry

Adam Cywinski, Paulina Bazgier
Ninety-one implantations of the Hanita Intensity, intraocular lens, including 34 patients in both eyes were performed over 18 months. In the majority (55 eyes), implantation was performed during refractive lens exchange, and in the remaining 36 cases, it was performed because of cataract. The qualifying examination, apart from the evaluation of standard parameters, included the evaluation of kappa and alpha angles, value of higher-order aberrations generated by the cornea, and pupil size. High postoperative satisfaction due to very good vision parameters (more than 80% of operated patients achieved vision to far distances ≥ 0.1 logMAR in the cataract group and 87% in the refractive lens exchange group, and almost 100% vision to near distances at the level of D-0.5 (Snellen chart) in the refractive lens exchange group) allows us to recommend this lens model in the most demanding group of patients qualified for refractive lens exchange. Statistical analysis showed that patients with the Hanita Intensity lens had significantly better visual acuity to far distances after surgery than those with the PanOptix (Alcon) lens. Conflict of Interest declaration. The authors declare that they have no affiliations with or involvement in any organization or entity with any financial interest in the subject matter or materials discussed in this manuscript.

Therapeutic Advances in Ophthalmology

Germán R. Bianchi
This ambispective, non-randomized study evaluated the visual performance, safety, and patient satisfaction of the Intensity SL (Hanita Lenses) five-focus pseudophakic intraocular lens (IOL) over a 24-month period. The study included 120 eyes of 60 patients who underwent femtosecond laser-assisted bilateral cataract surgery. At two years postoperatively, patients demonstrated excellent uncorrected visual acuity across far, intermediate, and near distances (binocular UDVA −0.05 ± 0.05 logMAR; UIVA 0.009 ± 0.05; UNVA 0.04 ± 0.06). Visual stability was maintained between the 6- and 24-month follow-ups. Dysphotopsias such as glare and halos were rare and mild. Patient satisfaction was remarkably high, with 95% reporting being "very satisfied" and 100% indicating they would choose the same IOL again. The findings support the long-term visual efficacy and quality-of-life benefits of the Intensity SL IOL.
Eye

Nature Eye

Jorge L. Alió et al.
This study assessed the refractive outcomes, optical performance, and quality of vision in patients implanted with the Intensity Hanita diffractive intraocular lens (IOL). The prospective, observational study included 64 eyes that underwent bilateral cataract surgery with Intensity IOL implantation, with evaluations conducted at 6 months post-surgery. Results indicated that 66% of eyes achieved uncorrected distance visual acuity (UDVA) of 0.10 logMAR or better. The defocus curve showed optimal vision at distance (0.02 ± 0.07 logMAR at 0.0 D) with stable performance through intermediate and near vision ranges (0.11 ± 0.08 logMAR at -1.5 D, 0.12 ± 0.12 logMAR at -2.5 D). Contrast sensitivity remained stable (p ≥ 0.06), and high-order aberrations were low, with a Strehl ratio of 0.12 ± 0.04. Mild halos and glare were reported but did not interfere significantly with daily activities, and the near activity visual questionnaire (NAVQ) indicated high patient satisfaction with near-vision tasks. These findings support the Intensity Hanita IOL’s effectiveness in providing clear vision across all distances through a five-foci distribution design that enhances depth of focus.

Fill in the details and receive the presentation